1{{Infobox concept1{{Infobox concept
2| name = Retention time2| name = Retention time
+3| image = chromatogram.svg
3| Symbol = t_R4| Symbol = t_R
4| Depends on = Column, mobile phase, gradient, flow, temperature5| Depends on = Column, mobile phase, gradient, flow, temperature
1819
19The strongest form combines retention matching with an orthogonal determination — [[Mass spectrometry|mass confirmation]] or a second chromatographic mode with different selectivity. This is what pharmacopoeial identification tests specify.{{r|usp621}}20The strongest form combines retention matching with an orthogonal determination — [[Mass spectrometry|mass confirmation]] or a second chromatographic mode with different selectivity. This is what pharmacopoeial identification tests specify.{{r|usp621}}
+21
+22== Drift and its causes ==
+23| Cause | Effect |
+24|---|---|
+25| Column ageing | Progressive shift, usually earlier |
+26| Temperature change | Shift in either direction |
+27| Mobile-phase preparation | Small shifts, run to run |
+28| Instrument dwell volume | Systematic offset between instruments |
+29| Gradient delay | Same, for gradient methods |
+30
+31Dwell-volume differences are the reason a method transferred between instruments often needs adjustment: the delay between the pump forming a gradient and it reaching the column differs, so retention shifts even with identical settings.{{r|usp621}}
2032