17Co-injection strengthens it substantially. Spiking the sample with the standard should enlarge the existing peak without producing a shoulder; a shoulder indicates two species.17Co-injection strengthens it substantially. Spiking the sample with the standard should enlarge the existing peak without producing a shoulder; a shoulder indicates two species.
1818
+19The strongest form combines retention matching with an orthogonal determination — [[Mass spectrometry|mass confirmation]] or a second chromatographic mode with different selectivity. This is what pharmacopoeial identification tests specify.{{r|usp621}}
+20
19== References ==21== References ==
20{{reflist}}22{{reflist}}