| 15 | Mass analysers differ in resolution and in mass accuracy. A quadrupole gives unit resolution and is adequate for confirming a mass to within a dalton; time-of-flight and orbital-trap analysers give sufficient resolution to determine a mass to a few parts per million, which distinguishes species a quadrupole cannot. | 15 | Mass analysers differ in resolution and in mass accuracy. A quadrupole gives unit resolution and is adequate for confirming a mass to within a dalton; time-of-flight and orbital-trap analysers give sufficient resolution to determine a mass to a few parts per million, which distinguishes species a quadrupole cannot. |
| 16 | 16 | ||
| + | 17 | The distinction matters for peptide work because the confounders are close in mass. [[Deamidation|Deamidation]] adds 0.984 Da, which unit resolution cannot resolve from the parent; oxidation adds 16 Da, which it can. A report stating an observed mass without stating the resolution of the instrument understates the ambiguity.{{r|usp1503}} | |