54Two observations follow. A tenfold error in assumed concentration produces a tenfold error in delivered mass with no change in the syringe reading and nothing visibly wrong. And at high concentrations a single graduation carries a large mass, so the resolution of the syringe becomes the limiting factor on dose precision: at 20 mg/mL, one unit is 200 µg, and half a graduation cannot be read reliably.{{r|iso7886b}}54Two observations follow. A tenfold error in assumed concentration produces a tenfold error in delivered mass with no change in the syringe reading and nothing visibly wrong. And at high concentrations a single graduation carries a large mass, so the resolution of the syringe becomes the limiting factor on dose precision: at 20 mg/mL, one unit is 200 µg, and half a graduation cannot be read reliably.{{r|iso7886b}}
5555
+56== Sources of error ==
+57; Scale confusion : Transferring a unit figure between U-100 and U-40 scales, as above.
+58; Unit-as-mass : Treating "units" as a quantity of drug. This is correct only for insulin, for which the international unit is defined by biological activity and for which the syringe scale was designed.{{r|who}}
+59; Graduation reading : Estimating between graduations. ISO 7886 permits a volume tolerance of a few percent of nominal capacity, which at the bottom of a 1 mL barrel is a substantial fraction of the reading.{{r|iso7886b}}
+60; Air and dead space : A bubble displaces solution; [[Syringe dead space|dead space]] retains it. Neither changes the barrel reading.
+61; Rounding : A computed 12.4 units cannot be drawn on a barrel graduated in whole units. Rounding to 12 is a 3% dose reduction, which matters more at low doses than high ones.
+62
56== References ==63== References ==
57{{reflist}}64{{reflist}}
59<ref name="iso7886b">ISO 7886-1, ''Sterile hypodermic syringes for single use — Part 1: Syringes for manual use''. International Organization for Standardization.</ref>66<ref name="iso7886b">ISO 7886-1, ''Sterile hypodermic syringes for single use — Part 1: Syringes for manual use''. International Organization for Standardization.</ref>
60<ref name="fdavet">US Food and Drug Administration, Center for Veterinary Medicine — guidance on U-40 insulin products and syringe compatibility.</ref>67<ref name="fdavet">US Food and Drug Administration, Center for Veterinary Medicine — guidance on U-40 insulin products and syringe compatibility.</ref>
+68<ref name="who">World Health Organization Expert Committee on Biological Standardization — definition of the international unit for insulin.</ref>