12Four choices in that description change the answer, and none of them is usually stated on a certificate:12Four choices in that description change the answer, and none of them is usually stated on a certificate:
1313
+14* '''The gradient.''' A shallow gradient resolves species that a steep one runs together. Two laboratories can obtain two honest and different figures for one vial by choosing differently here.
+15* '''The detection wavelength.''' At 214 nm the peptide bond dominates and most peptide-related impurities are seen. At 280 nm only aromatic residues respond, and a related substance without one is invisible.
+16* '''The integration rule.''' Valley-to-valley and drop-line integration assign the area under an unresolved shoulder differently, and the difference lands entirely on the principal peak.
+17* '''The integration threshold.''' Peaks below the threshold are not counted at all, so raising it raises the reported purity without any change to the material.{{r|usp1503}}