High-performance liquid chromatography (revision 15)
Old revision·02:58, 3 Jan 2025·SurmountSurma
| High-performance liquid chromatographySeparation technique | |
|---|---|
Detector response against time; peak area is proportional to the quantity of the eluting species. | |
| Abbreviation | HPLC |
| Separates by | Differential partition between stationary and mobile phases |
| Usual detector for peptides | UV absorbance at 214 nm |
| Key parameters | |
| Column | Chemistry, particle size, length, internal diameter |
| Mobile phase | Aqueous and organic components, modifier |
| Gradient | Composition change over time |
| Flow rate | Typically 0.2–1.5 mL/min |
| Analytical method infobox · conventions | |
High-performance liquid chromatography (HPLC) separates the components of a mixture by passing a solution through a column packed with fine particles. Components that interact more strongly with the packing move more slowly and emerge later; the detector records what emerges against time, producing a chromatogram.[1]
For peptides the dominant mode is reverse phase, in which the stationary phase is hydrophobic and the mobile phase is a water–acetonitrile mixture whose organic content is increased during the run. Detection is usually by ultraviolet absorbance at 214 nm, where the amide bond absorbs, so response is broadly proportional to the number of peptide bonds rather than to any particular side chain.[2]
Almost every number on a peptide certificate of analysis originates here. Understanding what the technique measures — and what it cannot measure — is therefore the single most useful piece of analytical background for reading such a document.[2]
How the separation works
[edit]A sample is injected into a stream of mobile phase and carried onto the column. Each component partitions continuously between the mobile phase, in which it moves, and the stationary phase, in which it does not. The fraction of time spent in each determines how long it takes to traverse the column — its retention time.[1]
In reverse-phase separation the stationary phase is a hydrocarbon chain, commonly C18, bonded to silica particles. Hydrophobic molecules are retained more strongly. Increasing the proportion of organic solvent in the mobile phase weakens that retention, so a gradient of increasing organic content elutes components in approximate order of hydrophobicity.
Resolution between two peaks depends on their retention difference, on peak width, and on the efficiency of the column. Efficiency improves with smaller particles and with longer columns, at the cost of higher back-pressure — the constraint that drove the development of sub-2-micron particles and the instruments capable of running them.[3]
What the chromatogram reports
[edit]The chromatogram is detector response against time. Peak area is proportional to the quantity of the eluting species multiplied by its response factor at the detection wavelength; peak position identifies it only by comparison with a known.[1]
| Quantity | Obtained from | Caveat |
|---|---|---|
| Area percent purity | Main peak area over total area | Normalised; independent of how much was injected |
| Identity | Retention matched to a Reference standard | Retention alone is weak identification |
| Content | Area against a calibrated standard | Requires a standard and a validated method |
| Impurity profile | Areas of minor peaks | Only species that elute and absorb are counted |
The last caveat is the one most often overlooked. Species that do not absorb at the detection wavelength — inorganic salts, many sugars, water — are invisible, and species that do not elute within the run remain on the column and are simply absent from the chromatogram. A clean trace is evidence about what was detected, not about what was present.[2]